Geometrical Product Specifications (GPS) - Acceptance and reverification tests for coordinate measuring machines (CMM) - Part 5: CMMs using multiple-stylus probing systems (ISO 10360-5:2000)
Geometrical Product Specifications (GPS) - Acceptance and reverification tests for coordinate measuring machines (CMM) - Part 5: CMMs using multiple-stylus probing systems (ISO 10360-5:2000)
Subscribe on standards with our subscription service. When you use our service you can be assured the latest editions and easy access.
Read more about SIS Subscriptions
This part of ISO 10360 specifies acceptance and periodic reverification tests of the performance of CMMs having
multiple-stylus probing systems, including systems with fixed multiple-styli attached to a single probe (e.g. "star" stylus), multiple-probing systems such as those with a stylus for each of their probes, and systems with articulating probing.
The acceptance test and reverification test specified in this part of ISO 10360 are applicable only to the following:
- CMMs using any type of contact-probing system;
- stylus system configurations able to be prequalified (i.e. where the repeatability of the system operation is sufficient so that a stylus qualification is not required for each use);
NOTE: In the case of prequalified configurations, a similar test could be devised for non-repeatable changes of stylus system configurations.
- spherical, hemispherical and disk-shaped styli.
These tests are not applicable to non-contact probing systems, which require special testing procedures outside the scope of this part of ISO 10360.